The detector is designed for daily in vivo dosimetry for megavoltage external beam radiation therapy. While other MOSFET detectors have been used clinically for surface measurements6–9 or temporarily placed into a body cavity during treatment,10–12 only the DVS detector is designed to be permanently implanted to measure dose over the entire course of treatment. The original prototype DVS detector (see Figure 1a) comprises a single MOSFET within the onboard circuit and an antenna for radiofrequency (RF) communication with an external hand-held reader. The MOSFET is the radio-sensitive element within the detector. Radiation damage increases the gate bias required to allow current to flow, and this increase is proportional to the dose reading. Thus, readings acquired just prior to and immediately following radiation are necessary to compute the measured dose. The detector is fully encapsulated in a biocompatible glass cylinder with a diameter of 3.25mm and a length of 25mm and is completely wireless. It can be inserted either peri-operatively or percutaneously with a cannula provided by the manufacturer. Because RF telemetry is used to detect voltage changes due to irradiation, these detectors do not require an internal power source. They are unbiased during irradiation, making these passive devices ideal for in vivo use. Unlike other detectors in common use, such as thermoluminescent detectors, photodiodes, and most other MOSFET detectors, the implantable DVS detector is factory-calibrated. (Another type of MOSFET detector, the OneDose,13 which is a single-use detector produced by Sicel Technologies and designed to measure skin dose, is also factory-calibrated.) Dose measurement performed with the RF reader can be completed immediately following treatment, and takes just seconds per detector. These characteristics greatly add to the utility and efficiency of the DVS.
In Vitro Studies—Prototype Detector
We have performed independent studies to evaluate and assess the performance of the MOSFET itself2 and the manufacturer’s calibration.5 Using an acrylic phantom (Quasar body phantom, Modus Medical Devices, London, Ontario, Canada) under 60Co irradiation, we found the MOSFETs showed a reproducibility to within ±2% in a fixed geometry and temperature-controlled environment.2 Using the factory calibration, we found the dose reproducibility to be within 0.4±2.1% (1σ) for the calibration dose of 2Gy delivered over 20 fractions. For a larger dose of 4Gy per fraction, the dose response was slightly overestimated (0.9±2.3% (1σ)), while for a smaller dose of 1Gy per fraction, the dose response was slightly underestimated (-3.3±3.6% (1σ)).5 We confirmed the detector’s visibility on CT images as well as its relative invisibility on megavoltage portal images. Due to its metal antenna, we expect it to flare on magnetic resonance images. We also suggested that sandwiching two MOSFETs in the circuit would improve the variation in measured dose response throughout the course of treatment.2 The manufacturer also performed studies of the detector’s performance.1,4 When sources of error including set-up, treatment planning, and machine output were considered, the expected performance of the detector was reported to be within ±5%.4 Angular variation was found to be within ±1.3% with respect to rotation about the detector’s long axis.1 Fading of the voltage reading due to electron-hole recombination14 was found to be less than 2% within 20 minutes.1 Temperature variations occurring between the pre- and post-irradiation voltage readings can have a significant effect on the performance of a MOSFET detector.15 We found the reproducibility of the detectors to be improved by about 3% when the temperature was well controlled.2 The manufacturer also noted the effect of temperature during in vitro measurements,1,4 and suggested this would not be an important issue for in vivo patient use,4 as the patient’s temperature should not change significantly during a single treatment fraction.
Clinical Studies—Prototype Detector
The first clinical studies were performed on 10 canine companion animals to monitor adverse events due to detector placement as well as detector migration following implantation.1 No adverse events due to implantation were reported. However, repeat CT scans showed migration of about 4mm in three of the dogs, and larger migrations of 35mm and 70mm in two others. This led to the development of a vicryl sock for surgical implantation and a vicryl plug for percutaneous implantation, both of which reduce the chance of migration during the healing process. Clinical studies in humans have also been reported by the manufacturer.3,4,16 To date, the original prototype has been successfully implanted into the breast, prostate, rectum, lung, and thigh for in vivo dosimetry during external beam radiation therapy. Of the first 18 patients,4 no unanticipated adverse events were reported due to implantation. All detectors performed successfully and only one detector migrated, which was placed in unconsolidated tissue in the lung. The authors reported that 12 of the 18 patients had dose deviations 5% or larger in at least 40% of all treatment sessions.4 Comparing these results with in vitro phantom measurements with an expected accuracy of ±5%, the authors suggested such deviations could have occurred due to discrepancies with treatment planning (e.g. not using heterogeneity corrections), errors in patient set-up, and patient motion. These results demonstrate the utility of true in vivo measurements with such detectors. The manufacturer has also briefly reported on a larger study involving 60 breast and prostate patients.16 Three cases of detector migration >5mm were reported, and again there were no significant adverse events due to implantation. The authors observed significant variation between the predicted and measured dose in at least 33% of the patients, supporting the results of the initial study. New ASIC Detector
The prototype DVS device has recently evolved to a new design (see Figure 1a), the ASIC, which includes a smaller glass capsule (2.1mm width, 20mm length) and a double-MOSFET circuit less sensitive to temperature fluctuations. Figure 1b shows the new computer-controlled RF reader. The detectors are factory-calibrated in commercial-sized lots and are intended for daily in vivo dosimetry at body temperature.
The manufacturer has specified the dose response to be within 5.5% (2σ) for doses up to 20Gy and 6.5% (2s) for doses up to 74Gy.17 We have performed in vitro measurements to investigate the new detector’s performance under 6-MV photon irradiation. Detectors were placed in a water tank at body temperature, and were irradiated to the calibration dose of 2Gy per fraction for 20 fractions. As shown in Figure 2, we have found the dose response to be well within the manufacturer’s specifications. Improvements to the RF reader design have made the system easier to use within the treatment room environment.
Conclusions
There is little doubt that the increasing complexity of radiation therapy also increases the cost of error. The development of a true implantable detector is a breakthrough in the field of in vivo dosimetry for radiation therapy, and the DVS manufactured by Sicel Technologies promises to measure dose within the treatment field or in the vicinity of normal tissues with a reasonable accuracy. In practice, we have found the system to be easy to use, permitting multiple detector readings within a matter of seconds. Due to fading of the voltage reading with time, measurements should be taken within 10 minutes of the completion of treatment, which should not be burdensome in a normal clinical environment. True in vivo dosimetry will allow confirmation of treatment delivery and better assessment of normal tissue toxicities. In vivo dosimetry can be used with kilovoltage imaging to provide complementary information, which may allow more accurate assessment for the cause of deviation from the treatment plan, including set-up error and patient motion.